CVE-2025-1204
固件中的"update"二进制文件存在隐藏功能漏洞(CWE-912),按下'C'按钮触发后尝试连接固定IP地址,绕过网络设置。攻击者若控制该IP可上传覆盖文件。建议移除Contec CMS8000设备或阻止相关IP地址。 2025-2-25 09:17:0 Author: claroty.com(查看原文) 阅读量:15 收藏

High Threat

CWE-912 HIDDEN FUNCTIONALITY:

The "update" binary in the firmware of the affected product sends attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function triggers if the 'C' button is pressed at a specific time during the boot process. If an attacker is able to control or impersonate this IP address, they could upload and overwrite files on the device.

Per FDA recommendation, CISA recommends users remove any Contec CMS8000 devices from their networks.

If asset owners cannot remove the devices from their networks, users should block 202.114.4.0/24 from their networks, or block 202.114.4.119 and 202.114.4.120.

Please note that this device may be re-labeled and sold by resellers.

Read more here: Do the CONTEC CMS8000 Patient Monitors Contain a Chinese Backdoor? The Reality is More Complicated….


文章来源: https://claroty.com/team82/disclosure-dashboard/cve-2025-1204
如有侵权请联系:admin#unsafe.sh